The following is the specific cleaning process of ceramic membrane device:
Pure water cleaning
Flush the equipment to the system with pure water.
Alkaline cleaning
Using pure water as 1% to 3%, the temperature of 60-70 ℃ of NaOH aqueous solution, cleaning about 50 minutes. Rinse with pure water to neutral.
Acid cleaning
Using pure water as 1%, citric acid aqueous solution for 30 to 40 ℃, temperature cleaning about 50 minutes. Rinse with pure water to neutral.
Note:
For the first time, the film should be recycled. The method is:
Alkaline cleaning:
With temperature is about 60-70 ℃, the concentration of 1% NaOH solution cleaning, time for about 30 minutes with deionized water rinse to neutral;
Pickling:
With temperature is about 30 to 40 ℃, the concentration of 1% lemon acid solution to clean, time for about 30 minutes with deionized water rinse to neutral.